MRI Safety in Clinical Practice
Corresponding author: [email protected]
Abstract
Magnetic resonance imaging uses no ionising radiation and was long regarded for that reason as an intrinsically safe examination, but an MR system combines a static field of 1.5–3 T that never switches off, rapidly switched gradient fields and a radiofrequency field at the Larmor frequency, together with a large store of liquid helium and, in many examinations, an intravenous gadolinium chelate. Each carries its own mechanism of harm, and because none leaves a cumulative dose, MR safety rests on access control, screening and the management of exposure during the scan rather than on dose limits and personnel monitoring. The discipline was shaped by accidents, including the death of a six-year-old boy struck by a steel oxygen cylinder in 2001. This chapter covers the sources of hazard, the static field, gradient and radiofrequency fields, the four-zone model and MR personnel, patient screening, implants, devices and foreign bodies, gadolinium-based contrast agents, pregnancy and special patient groups, and emergencies in Zone IV.
Author Affiliations
- 1 Scholar, Saraswati College of Pharmacy, SGC Group, Gharuan 140413, Mohali, Punjab, India.